Prior-image-based MRI reconstruction with global variation regularization
编号:90 访问权限:仅限参会人 更新:2021-10-30 22:03:44 浏览:670次 张贴报告

报告开始:2021年11月13日 09:25(Asia/Shanghai)

报告时间:5min

所在会场:[Pos] Poster [Pos] Poster Session

视频 无权播放 演示文件

提示:该报告下的文件权限为仅限参会人,您尚未登录,暂时无法查看。

摘要
Purpose: 
Reconstructing MR images from under-sampling kspace is a challenge due to incomplete information. In compressed sensing (CS) reconstruction, sparsity transform, such as total variation (TV) and wavelet transform, was proposed to compensate for the missing information by the assumption of the sparsity property of the natural image. If there exist MR images with different contrasts (T1, T2) or images from other models (such as CT, PET), more prior information can be achieved. The images are usually accessible in clinical practice, such as radiotherapy. The structure-guided total variation (sTV) has been proposed to utilize similar structure information between different images. The study aims to further exploit the prior information from the existing image to assist image reconstruction. The CS algorithm is to solve a minimization problem. The TV function constraints the image to be smooth. While there usually exist some regions, in which the signal is almost uniform, such as air, these regions can be extracted from the prior image, then utilized in the regularization. Here, we propose the global variation (GV) function.
Methods:
A numerical simulation experiment was designed to examine the performance of the proposed GV. A pair of aligned T1-weighted and T2-weighted MR images were used as the known prior image and the ground truth image for reconstruction. The under-sampled kspace was got by a masked Fourier transform. The under-sampling rate is 0.13, GV mask was extracted from the reference by selecting the signal value smaller than 5% of the maximal value of the image. Five regularizations were tested: no prior, TV, sTV, GV, sTV+GV. The regularization parameters were selected experientially as 0.03. The minimization function is solved by a nonlinear conjugate function algorithm.
Results and discussions:
With an under-sampling rate of 0.13, the reconstructed image without a regularization term (i.e., no prior) shows strong aliasing artifacts. Lots of them are located out of the head region, referring to signal leakage. TV regularization reduces the artifacts and sTV regularization goes further, but there still exists lots of leaked signals. GV regularization prevents signal leakage, leading to a cleaner out-head region. Compared with no prior, the MSE within the head is also reduced. Combining GV with sTV, the artifacts are further suppressed.
Conclusions: 
The region with a uniform signal extracted from the prior image contributes to the under-sampled image reconstruction. The GV regularization is capable to utilize this prior information to suppress the aliasing artifacts.
关键词
fast MRI, regularization, prior image
报告人
飞 王
北京大学肿瘤医院

稿件作者
飞 王 北京大学肿瘤医院
莎 李 北京大学肿瘤医院
艺宝 张 北京大学肿瘤医院
昊 吴 北京大学肿瘤医院
发表评论
验证码 看不清楚,更换一张
全部评论
重要日期
  • 会议日期

    11月13日

    2021

    11月14日

    2021

  • 09月30日 2021

    报告提交截止日期

  • 11月14日 2021

    注册截止日期

主办单位
IEEE北京分会
中国生物医学工程学会医学物理分会
中国电子学会生命电子学分会
承办单位
中国科学技术大学
安徽省生物医学工程学会
联系方式
移动端
在手机上打开
小程序
打开微信小程序
客服
扫码或点此咨询