Reliable Calibration of Cone Beam CT Using Size-coding Markers
编号:74 访问权限:仅限参会人 更新:2021-10-30 18:53:13 浏览:596次 张贴报告

报告开始:2021年11月14日 22:55(Asia/Shanghai)

报告时间:25min

所在会场:[PS1] Plenary Session 1 [CT1] Workshop on CT

视频 无权播放 演示文件

提示:该报告下的文件权限为仅限参会人,您尚未登录,暂时无法查看。

摘要
Geometric calibration is essential to counter the gantry imperfection. It is a process of projecting 3D markers to 2D and find the corresponding mapping matrices. Therefore, to achieve accurate calibration, the same marker in both 2D and 3D must be identified. However, in early reported phantom designs, most of the marker are of the same sizes. It makes the identification of the markers difficult. We proposed a size-coding approach that the markers of varying sizes are embedded into the phantom to make the marker string unique and more immune from the artifacts, such as the overlap of the markers in projection images. The experimental results showed the proposed method is simpler and effective to achieve accurate calibration that enables the cone beam CT reconstruction.
关键词
geometric calibration, size-coding markers, projection matrix
报告人
Dan Wang
首都医科大学

稿件作者
Dan Wang Capital Medical University
Yunxian Zhang Capital Medical University
Yumeng Zhang Capital Medical University
Fan Peng Capital Medical University
Shangqi Cui Capital Medical University
Da He Beijing Jishuitan Hospital
Bo Liu Bejing Jishuitan Hospital
Zhi Yang Capital Medical University
发表评论
验证码 看不清楚,更换一张
全部评论
重要日期
  • 会议日期

    11月13日

    2021

    11月14日

    2021

  • 09月30日 2021

    报告提交截止日期

  • 11月14日 2021

    注册截止日期

主办单位
IEEE北京分会
中国生物医学工程学会医学物理分会
中国电子学会生命电子学分会
承办单位
中国科学技术大学
安徽省生物医学工程学会
联系方式
移动端
在手机上打开
小程序
打开微信小程序
客服
扫码或点此咨询